J Arthroplasty. 2016 Jan; 31(1): 162–167.

Long-Term Results of Total Hip Arthroplasty with 28 millimeter Cobalt-Chromium Femoral Heads on Highly Cross-linked Polyethylene in Patients 50 years and Less

Jeffrey B. Stambough, MD,1 Gail Pashos, BS MT,1 Frank C. Bohnenkamp, MD,2 William J. Maloney, MD,3 John M. Martell, MD,4 and John C. Clohisy, MD1
Hip

Highly cross-linked polyethylene (HXLPE) is the most commonly used bearing surface in total hip arthroplasty (THA) because of its superior wear properties, but long-term results in young patients are limited. We report on the clinical outcome, radiographic wear patterns and survivorship of 72 patients ≤50 years old who had a 28-millimeter cobalt-chromium femoral head on HXLPE acetabular liner. Mean and median true linear wear rates at average ten-year follow-up were 0.0104 and 0.016 mm per year +/− 0.07 mm. Mean and median two-dimensional volumetric wear rates were 12.79 mm3 and 5.834 mm3 per year +/− 26.1mm3 as determined by Martell analysis. As a result of the minimal wear profile, there was no evidence of radiographic osteolysis and no wear-related revisions.