The Journal of Arthroplasty, Volume 27, Issue 7, 1413.e5 - 1413.e9

Histologic Retrieval Analysis of a Porous Tantalum Metal Implant in an Infected Primary Total Knee Arthroplasty

Sambaziotis, Chris et al.
Knee

Porous tantalum (Zimmer, Inc, Warsaw, Ind) has the theoretical advantage of improved biologic fixation because of its high porosity, interconnected pore space, and modulus of elasticity. We present a case report documenting the retrieval and bone ingrowth analysis of a porous tantalum tibial component in an infected total knee arthroplasty. Results demonstrated a significantly larger amount of bone ingrowth present in the tibial posts (36.7%) when compared with the bone ingrowth into the tibial baseplate (4.9%) (P < .001). The data suggest that bone ingrowth seen in the plugs as well as baseplate was suggestive of viable bone tissue with healthy bone marrow, osteocytes, and lamella, resulting in a well-fixed tibial implant even at revision surgery for an infected total knee arthroplasty.


Link to article