The Journal of Arthroplasty, ISSN: 0883-5403, Vol: 24, Issue: 5, Page: 806-814

Fixed-Angle Screws vs Standard Screws in Acetabular Prosthesis Fixation: A Cadaveric Biomechanical Study

Hugate, Ronald R; Dickey, Ian D; Chen, Qingshan; Wood, Christina M; Sim, Franklin H; Rock, Michael G
Hip

Secure fixation of acetabular components in total hip arthroplasty can be challenging. The purpose of this study was to perform biomechanical analysis of cup fixation strength using fixed-angle vs standard screw fixation. Multihole, porous-backed acetabular prostheses were implanted in both acetabuli of 8 cadaveric pelves using standard press-fit techniques. Fixed-angle screws were used on the left side, and standard cancellous screws were used in the right. The use of fixed-angle screws enhanced acetabular fixation substantially under subfailure cyclic loading conditions and load-to-failure. The triradiate screw configuration increases the bending moment required to fail the specimens as well. Fixed-angle screws may be useful for achieving rigid fixation of acetabular prostheses in challenging clinical scenarios.