The Journal of Arthroplasty, ISSN: 0883-5403, Vol: 25, Issue: 7, Page: 1169.e5-1169.e7

Failure of the Modular Neck in a Total Hip Arthroplasty

Chris J. Dangles; Carl J. Altstetter
Hip
Modular femoral necks in total hip arthroplasty are promoted as giving the surgeon more options during surgery and as being safe. We report a case of fatigue fracture in the modular neck of a femoral implant 3 1/2 years after implantation. The probable design errors leading to product failure are discussed.

Link to article