Bone Joint J 2021;103-B(6 Supple A):32–37.

Excellent mid-term follow-up for a new 3D-printed cementless total knee arthroplasty

Santiago Restrepo, Eric B. Smith, William James Hozack
Knee

Aims

Cementless total knee arthroplasty (TKA) offers the potential for strong biological fixation compared with cemented TKA where fixation is achieved by the mechanical integration of the cement. Few mid-term results are available for newer cementless TKA designs, which have used additive manufacturing (3D printing). The aim of this study was to present mid-term clinical outcomes and implant survivorship of the cementless Stryker Triathlon Tritanium TKA.

Methods

This was a single institution registry review of prospectively gathered data from 341 cementless Triathlon Tritanium TKAs at four to 6.8 years follow-up. Outcomes were determined by comparing pre- and postoperative Knee Injury and Osteoarthritis Outcome Score for Joint Replacement (KOOS JR) scores, and pre- and postoperative 12-item Veterans RAND/Short Form Health Survey (VR/SF-12) scores. Aseptic loosening and revision for any reason were the endpoints which were used to determine survivorship at five years.

Results

At mid-term follow-up, the mean KOOS JR score improved significantly from 33.14 (0 t0 85, standard deviation (SD) 21.88) preoperatively to 84.12 (15.94 to 100, SD 20.51) postoperatively (p < 0.001), the mean VR/SF-12 scores improved significantly from physical health (PH), 31.21 (SD 5.32; 23.99 to 56.77) preoperatively to 42.62 (SD 10.72; 19.38 to 56.82) postoperatively (p < 0.001) and the mental health (MH), 38.15 (SD 8.17; 19.06 to 60.75) preoperatively to 55.09 (SD 9.64; 19.06 to 66.98) postoperatively (p < 0.001). A total of 11 revisions were undertaken, with an overall revision rate of 2.94%, including five for periprosthetic joint infection (1.34%), three for loosening (0.80%), two for instability (0.53%), and one for pain (0.27%). The overall survivorship was 97.06% and survivorship for aseptic loosening as the endpoint was 98.40%, with a 99.5% survivorship of the 3D-printed tibial component.

Conclusion

This 3D-printed cementless total knee system shows excellent survivorship at mid-term follow-up. This design and the ability to obtain cementless fixation offers promise for excellent long-term durability.


Link to article