The Journal of Arthroplasty, Volume 33, Issue 1, 268 - 272

Does the Additional Articulation in Retrieved Bipolar Hemiarthroplasty Implants Decrease Trunnionosis Compared to Total Hip Arthroplasty?

Del Balso, Christopher et al.
Hip

Background

Trunnionosis at the modular head-neck taper interface in metal-on-polyethylene total hip arthroplasty (MoP THA) has been shown to occur, and represents a potential mode of MoP THA failure. The purpose of the present investigation is to elucidate differences in fretting and corrosion at the head-neck taper interface of prostheses retrieved from bipolar hemiarthroplasty (BH) and MoP THA.

Methods

A retrieval analysis of BH and MoP THA prostheses featuring a single taper design from a single manufacturer and in vivo for a minimum 2 years was performed. Fifteen femoral heads of 28-mm diameter and corresponding femoral stems retrieved from BH were compared with MoP THA implants matched based on time in vivo and head length (28 mm, −3 mm to 28 mm, +8 mm). Fretting and corrosion damage scoring was completed under stereomicroscopic visualization.

Results

Femoral head bore tapers retrieved from BH exhibited decreased overall fretting (P = .02), when compared to those retrieved from MoP THA. Total corrosion scores for all retrieved implants were positively correlated with implantation time (ρ = 0.54, P < .02).

Conclusion

Femoral heads retrieved from BH exhibit decreased fretting damage compared to those retrieved from MoP THA. The added articulation in BH implants may decrease torque produced at the head-neck taper junction, thereby decreasing fretting. Increased fretting damage in implants from MoP THA is not associated with increased corrosion in 28-mm heads of this taper design. The longer a BH or MoP THA prosthesis is implanted, the greater the risk of damage due to corrosion.