Customized implants for acetabular Paprosky III defects may be positioned with high accuracy in revision hip arthroplasty. International Orthopaedics (SICOT) 43, 2235–2243 (2019) doi:10.1007/s00264-018-4193-3

Customized implants for acetabular Paprosky III defects may be positioned with high accuracy in revision hip arthroplasty

Weber, M., Witzmann, L., Wieding, J. et al.
Hip

Purpose

In revision hip arthroplasty, custom-made implants are one option in patients with acetabular Paprosky III defects.

Methods

In a retrospective analysis, we identified 11 patients undergoing cup revision using a custom-made implant. The accuracy of the intended position of the implant was assessed on post-operative 3D CT and compared to the pre-operative 3D planning in terms of inclination, anteversion, and centre of rotation. In addition, the accuracy of post-operative plain radiographs for measuring implant position was evaluated in relation to the 3D CT standard.

Results

We found a mean deviation between the planned and the final position of the custom-made acetabular implant on 3D CT of 3.6° ± 2.8° for inclination and of − 1.2° ± 7.0° for anteversion, respectively. Restoration of center of rotation succeeded with an accuracy of 0.3 mm ± 3.9 mm in the mediolateral (x) direction, − 1.1 mm ± 3.8 mm in the anteroposterior (y) direction, and 0.4 mm ± 3.2 mm in the craniocaudal (z) direction. The accuracy of the post-operative plain radiographs in measuring the position of the custom-made implant in relation to 3D CT was 1.1° ± 1.7° for implant inclination, − 2.6° ± 1.3° for anteversion and 1.3 mm ± 3.5 mm in the x-direction, and − 0.9 mm ± 3.8 mm in the z-direction for centre of rotation.

Conclusion

Custom-made acetabular implants can be positioned with good accuracy in Paprosky III defects according to the pre-operative planning. Plain radiographs are adequate for assessing implant position in routine follow-up.


Link to article