The Journal of Arthroplasty, ISSN: 0883-5403, Vol: 23, Issue: 1, Page: 26-29

Custom-Fit Total Knee Arthroplasty (OtisKnee) Results in Malalignment

Klatt, Brian A; Goyal, Nitin; Austin, Matthew S; Hozack, William J
Knee

We report on 4 patients who underwent total knee arthroplasty with OtisKnee system (OtisMed, Hayward, Calif). An image-free computer navigation system was used to evaluate the deformities and the recommended cuts. The recommended custom cuts were as follows: valgus/varus cuts on the femur (5.5° valgus to 0.5° varus) in reference to the mechanical axis, flexion cuts on the femur (4°-9° of flexion); femoral rotation was within 1° of the epicondylar axis; valgus/varus cut on the tibia (3° of valgus to 7.5° of varus); tibial slope (5.5° of anterior slope to 0.5° of posterior slope). The custom OtisKnee system guides recommended alignment of the components that was more than 3° off of mechanical axis. The potential for malalignment with this system places implants at high risk of early failure.


Link to article