Knee Surgery, Sports Traumatology, Arthroscopy January 2016, Volume 24, Issue 1, pp 64–73

Computer-assisted navigation is beneficial both in primary and revision surgery with modular rotating-hinge knee arthroplasty

Ochs, B.G., Schreiner, A.J., de Zwart, P.M. et al.
Knee

Purpose

The objective of the present study was to explore the effect of navigation on the reconstruction of the mechanical leg axis, implant positioning and the restoration of the joint line in hinged knee arthroplasty in vivo. We present the first 1- to 3-year clinical and radiological results following computer-navigated implantation of the EnduRo modular rotating-hinge knee arthroplasty system (Aesculap AG, Tuttlingen, Germany) as a primary or revision implant.

 

Methods

Thirty-one patients were analysed retrospectively. Indication was revision surgery in 18 patients and complex primary surgery in 13. The clinical and radiological results of 31 patients with a minimum follow-up of 12 months (mean 22.2 ± 6.2 months) were recorded. Age at follow-up was 55.2 ± 9.9 years.

 

Results

The absolute varus-valgus deviation from the neutral mechanical leg axis was determined at 5.1° ± 5.1° preoperatively and 2.1° ± 1.4° postoperatively. No intraoperative complications or problems with the navigation system were observed. At latest follow-up, no component loosening was detected. Based on the Knee Society Score, a knee score of 64.9 ± 17.7 points and a function score of 67.2 ± 27.3 points were achieved.

 

Conclusions

Encouraging short-term clinical and radiological results with the computer-navigated implantation of the modular rotating-hinge EnduRo knee arthroplasty system were found in both primary and revision surgery. The navigation facilitated the reconstruction of the leg axis, implant positioning and the restoration of the joint line.

Level of evidence

IV.


Link to article