Acta Orthopaedica, 86:4, 440-443

Computed tomography scanogram compared to long leg radiograph for determining axial knee alignment

Thomas J Holme, Johann Henckel, Kai Hartshorn, Justin P Cobb & Alister J Hart
Knee

Background and purpose — Supine computed tomography scanogram (CTS) is a commonly used alternative to weight bearing long leg plain radiograph (LLR) in measuring knee alignment. No published studies have validated its use in the native knee and the post-unicompartmental replacement knee (UKR). We quantified the difference in measurements obtained from CTS and LLR for knee alignment.

 

Patients and methods — Supine CT scanograms and weight bearing long leg plain anteroposterior radiographs were obtained for 40 knees (in 25 patients), 17 of which were native, and 23 of which were post-UKR. The mechanical and anatomical axes of the tibio-femoral joint were measured. Bland-Altman plots were used to calculate the 1.96 standard deviation limits of agreement between CTS and LLR. Intraclass correlation was used to assess intra-rater and inter-rater reliability (where values > 0.81 indicate very good reliability).

 

Results — CTS and LLR were equally reliable in measurement of the mechanical and anatomical axes of the tibio-femoral joint (intraclass correlation coefficient (ICC) > 0.9 for all parameters). Statistically significant and clinically relevant differences were found between CTS and LLR in measurement of the mechanical axis (limits of agreement: UKR −3.2° to 6.3°; native −3.2° to 5.6°) and the anatomical axis (limits of agreement: UKR −3.7° to 8.7°; native −2.0° to 8.8°).

 

Interpretation — Although it is a reliable tool, CTS is not necessarily an accurate one for measurement of knee alignment when compared to LLR. We recommend that CTS should not be used as a substitute for LLR in measurement of the mechanical or anatomical axes of the knee.