JBJS, April 1, 2006, Volume 88, Issue 4

Ceramic Failure After Total Hip Arthroplasty with an Alumina-on-Alumina Bearing

Youn-Soo Park, MD Sung-Kwan Hwang, MD Won-Sik Choy, MD Yong-Sik Kim, MD Young-Wan Moon, MD Seung-Jae Lim, MD
Hip
Background: The mechanical properties of alumina ceramic, now in its third generation, have been markedly improved through the evolution of design features and manufacturing processes and the introduction of proof-testing. Nonetheless, because of the lack of ductility of alumina ceramic, there is concern regarding the risk of fracture during insertion or in vivo use. The purpose of the present study was to present a multicenter review of primary total hip arthroplasties performed with use of a polyethylene-ceramic composite liner combined with a ceramic femoral head, with particular attention to failure of the ceramic bearing.
Methods: We evaluated 357 primary total hip arthroplasties that had been performed in 319 patients with use of a contemporary alumina-on-alumina bearing design incorporating a polyethylene-ceramic composite liner within a titanium-alloy shell coupled with a 28-mm-diameter ceramic femoral head. The procedures were performed at four participating centers from 1998 to 2001. Ceramic failure without trauma occurred in six hips (1.7%). All of these hips were revised, and the retrieved alumina implants were examined by means of visual inspection and scanning electron microscopy equipped with energy-dispersive x-ray spectrometry.
Results: Two femoral heads fractured during the first postoperative year, and four alumina liners fractured after an average of 36.8 months in vivo. All four of the explanted alumina liners revealed evidence of rim contact with the metal neck of the femoral component. Composition analysis confirmed that surface-stain materials were titanium particles transferred from the femoral component.
Conclusions: Despite the theoretical improvement in the fracture toughness of a polyethylene-alumina composite liner, a relatively high rate of catastrophic ceramic bearing surface failure was still observed at the time of short-term follow-up. This finding prompted us to discontinue the use of this type of alumina bearing design.
Level of Evidence: Therapeutic Level IV. See Instructions to Authors for a complete description of levels of evidence.