The Journal of Arthroplasty, ISSN: 0883-5403, Vol: 22, Issue: 3, Page: 369-376

Cementless Hemispheric Hydroxyapatite-Coated Sockets for Acetabular Revision

Geerdink, Carel H; Schaafsma, Jelle; Meyers, Will G M; Grimm, Bernd; Tonino, Alphons J
Hip
This study describes our experience with a hydroxyapatite-coated uncemented hemispherical component used for revision of 72 patients with aseptic loosening of the acetabular component. Preoperative Paprosky classification of the acetabular defects was 1 type I, 35 type II, and 36 type III; according to American Academy of Orthopaedic Surgeons, 14 segmental, 10 cavitary, and 48 combined. The mean follow-up was 7.6 years (range, 5.0-13.0). Complications were seen in 7 cases (9.6%). Seventy acetabular components (97.2%) showed bone ongrowth on the radiographs. The survival rate was 90.8% after a mean follow-up of 7.6 years (range, 5.0-13.0) when revision for any reason is the end point (confidence interval, 80.5%-100%) and 98.1% (confidence interval, 94.5%-100%) when aseptic loosening of the cup is the end point. We conclude that hydroxyapatite-coated cups for acetabular revision show promising results.

Link to article