Cement mantle thickness does not influence serum or local gentamicin concentrations in hybrid total hip arthroplasty: a randomised controlled trial
van IJperen, W., Van Quickenborne, D., Buyl, R., & Scheerlinck, T.Hip
Introduction:
We investigated the in vivo gentamicin elution kinetics of Hi-Fatigue Gentamicin Bone Cement (AAP Biomaterials GmbH) in serum and drain fluid after hybrid hip arthroplasty and the relationship with cement mantle thickness.
Methods:
We compared in a randomised, non-blinded prospective study, the local and systemic gentamicin concentrations in 2 groups. The thin cement mantle group (n = 16) received a stem implanted line-to-line with the broach, whereas the thick group (n = 14) had an undersized stem. Gentamicin concentrations were measured in drain fluid and serum at set intervals for 3 days postoperatively.
Results:
In both groups, local gentamicin concentrations were similar. After a high initial burst above the minimal inhibitory concentration (thin: 57.2 mg/L (SD 34.4), thick: 54.9 mg/L (SD 19.9), p = 0.823) local gentamicin concentrations declined rapidly. In both groups, serum concentrations never exceeded toxic levels (maximum 1.08 mg/L).
Conclusion:
In hybrid total hip arthroplasty, Hi-Fatigue Gentamicin Bone Cement resulted in effective and safe gentamicin concentrations.
Link to article