The Journal of Arthroplasty, ISSN: 0883-5403, Vol: 26, Issue: 6, Page: 976.e1-976.e5

Catastrophic Failure of a Metal-on-Metal Total Hip Arthroplasty Secondary to Metal Inlay Dissociation

Michael J. Alaia; Alan J. Dayan
Hip

Metal-on-metal bearing surfaces in total hip arthroplasty have been recently shown to have acceptable survivorship properties (J Bone Joint Surg Am. 2006;88:1183; J Bone Joint Surg Am. 2006;88:1173), and they have certain advantages and disadvantages when compared to conventional metal-on-polyethylene bearing surfaces. Like traditional metal-on-polyethylene bearings, these metal-on-metal implants may also suffer from catastrophic failure. This case report represents an unusual situation in a 57-year-old man in which dissociation of a metal inlay in a metal-on-metal total hip arthroplasty resulted in articulation of the inferior aspect of the inlay with the femoral neck, leading to femoral neck notching, extensive periprosthetic soft tissue metallosis, osteolysis, and subsequent prosthetic catastrophic failure.


Link to article