The Journal of Arthroplasty, Volume 29, Issue 9, 1871 - 1876

Can Post-Cam Function Be Replaced by Addition of a Third Condyle in PS TKA?

Nakamura, Shinichiro et al.
Knee

The purpose of the study was to analyze the effectiveness of an additional ball and socket articulation in implanted knees and whether it can replicate post-cam function. Fifteen knees implanted with a cruciate substituting (CS) polyethylene without a post and ten knees implanted with a posterior stabilized (PS) polyethylene with a post were analyzed using 3D model fitting approach. Two types of designs showed similar posterior translation and similar axial rotation. Most of the contact points at the ball and socket joint stayed within the socket height for the PS group. This study indicates that the ball and socket joint is able to function as a replacement of the post-cam mechanism, which might serve as a new way to achieve posterior stability.


Link to article