Clinical Orthopaedics and Related Research: November 2004 - Volume 428 - Issue - p 198-206

Backside Wear of Miller-Galante I and Insall-Burstein II Tibial Inserts

Taki, Naoya MD*; Goldberg, Victor M MD*†; Kraay, Matthew J MD*†; Rimnac, Clare M PHD*‡
Knee

The objective of this study was to determine if there was a difference in the amount, type, and location of backside wear in the Miller-Galante I and Insall-Burstein II PE tibial inserts. A secondary objective was to determine if backside wear damage in these two designs was a function of clinical factors (patient height, weight, gender, age, and length of time of implantation), shelf life of the PE tibial insert, and tibial component thickness. Backside wear damage was assessed on 24 Miller-Galante I and 11 Insall-Burstein II tibial inserts (implantation time, 0.5–12.4 years). For both groups combined, implantation time was positively correlated to wear damage and to PE peg height into screw holes. The Miller-Galante I group had significantly larger PE pegs than the Insall-Burstein II group. The Miller-Galante I group had significantly more burnishing and larger PE pegs posteriorly than anteriorly. There was no correlation between insert shelf life before initial surgery and backside wear. The thinner the component, the larger the total damage scores in the Miller-Galante I group. This study supports the hypothesis that backside wear of PE tibial inserts may be influenced by design and component thickness and by clinical factors.