The Journal of Arthroplasty, Volume 30, Issue 12, 2354 - 2359

A Novel Method for Assessment of Polyethylene Liner Wear in Radiopaque Tantalum Acetabular Cups: Clinical Validation in Patients Enrolled in a Randomized Controlled Trial

Troelsen, Anders et al.
Hip

Conventional radiostereometric analysis (RSA) for wear is not possible in patients with tantalum cups. We propose a novel method for wear analysis in tantalum cups. Wear was assessed by gold standard RSA and the novel method in total hip arthroplasty patients enrolled in a randomized controlled trial receiving either titanium or tantalum cups (n = 46). The novel method estimated the center of the head using a model based on identification of two proximal markers on the stem and knowledge of the stem/head configuration. The novel method was able to demonstrate a pattern of wear that was similar to the gold standard in titanium cups. The novel method offered accurate assessment and is a viable solution for assessment of wear in studies with tantalum cups.