International Orthopaedics March 2014, Volume 38, Issue 3, pp 503–508

A comparison of conventional and patient-specific instruments in total knee arthroplasty

Daniilidis, K. & Tibesku, C.O.
Knee

Purpose

Several authors have observed that standard instrumentation (SI) may be insufficient for addressing component malalignment. Patient-matched cutting blocks (PMCB) technology was introduced to improve surgeons’ ability to achieve a neutral postoperative mechanical axis following total knee arthroplasty (TKA). The current retrospective study was designed to compare the ability of SI and PMCB to achieve a hip-knee-ankle angle (HKA) within ±3° of the ideal alignment of 180°.

Methods

Between October 2009 and December 2012, 170 TKAs in 166 patients (four bilateral) using VISIONAIRE (Smith & Nephew) PMCB technology were performed. Additionally, 160 TKAs in 160 consecutive patients that had received a total knee arthroplasty using SI during the same time period were used as a control group, All surgeries were performed by the same surgeon. Standardized pre- and postoperative long-leg standing x-rays were retrospectively evaluated to compare the two patient cohorts.

Results

X-rays were available for analysis for 156 knees in the SI group and 150 in the PMCB group. The average post-surgical HKA was 178.7 ± 2.5 in the SI group and 178.4 ± 1.5 in the PMCB group. However, the rate of ± 3° outliers was 21.2 % in the SI group and 9.3 % in the PMCB group. There were no intraoperative complications with the use of PMCB technology or SI.

Conclusions

PMCB technology proved superior to conventional instrumentation in achieving a neutral mechanical axis following TKA. Further follow-up will be needed to ascertain the long-term impact of these findings.