International Orthopaedics April 2014, Volume 38, Issue 4, pp 703–709

A cementless hip system with a new surface for osseous integration

Lass, R., Kolb, A., Skrbensky, G. et al.
Hip

Purpose

The failure of total hip systems caused by wear-particle-induced loosening has focused interest on factors potentially affecting wear rate. Remnants of the blasting material were reported on grit-blasted surfaces for cementless fixation. These particles are believed to cause third-body wear and implant loosening. The purpose of this study was to evaluate the early clinical and radiological outcomes of a cementless hip system with a new, contamination-free, roughened surface with regard to prosthesis-related failures.

Methods

Between May 2004 and March 2009, 202 consecutive primary total hip arthroplasties (THAs) (192 patients with a mean age of 62.6 years) were performed using a cementless stem (Hipstar®) and a hemispherical acetabular cup (Trident®).

Results

At a minimum follow-up of two years, five revisions (2.5 %) due to aseptic loosening of the stem and three (1.5 %) of the cup were necessary. The cumulative rate of prostheses survival, counting revision of both components and with aseptic failure as end point, was 92.9 % at 8.8 years. Radiolucent lines up to three millimetres were evaluated in the proximal part of the femur in 61 % of cases.

Conclusions

Although the incidence of radiolucent lines was decreased, the revision rate was considerably increased compared to other uncemented hip implants with grit-blasted surfaces in the short- to mid-term follow-up of our study. Subsequent studies are needed to confirm whether these changes in implant material and surface affect the radiological and clinical outcome in the long term.